Settings/Performance
Intended use: Patients >10 kg
Ventilation Modes:
- PSV
- PSV(T)
- PCV
- PCV(Tgv)
- PCV(A)
- PCV(A+Tgv)
- VCV
- VCV(A)
- CPAP
Device Mode: Home, Clinical
Inspiratory Pressure: 4 to 50 cmH2O
PEEP:
- 2 cmH2O (leakage circuit),
- 0 cmH2O (circuit with active exhalation valve) to 30 cmH2O,
- Pressure – 2 cmH2O or Min Pressure -2 cmH2O
Breath Rate (PCV, VCV): 4 to 40 bpm
Inspiratory Time: 0.3 to 5 s
Backup Inspiratory Time: 0.3 to 5 s (PSV)
Rise Time:
- 1 to 9 (PSV & PCV)
- 50 – 90 %, Off (VCV)
Inspiratory Trigger:
- 1 to 9 (PSV, PCV & VCV)
- Off (PCV & VCV)
Expiratory Trigger: 1 to 9 (PSV)
Minimum Inspiratory Time: Off, 0.3 to 3 s
Maximum Inspiratory Time: 0.3 to 3 s,
Off Backup Rate (PSV): 4 to 40 bpm
Target Volume: 100 to 2500 ml
Maximum Pressure: 4 to 50 cmH2O
Minimum Pressure: 4 cmH2O to 50 cmH2O
Tidal Volume: 100 to 2500 ml
Flow Pattern: Square, decelerating
CPAP: 4 to 20 cmH2O
Power Supplies
Mains supply: 100 to 240 V AC
External battery: 24 V DC
Click-on battery: 8 hours
Internal battery: 4 hours
Main Alarms
High Pressure, Low Pressure, High PEEP, Low PEEP, High Vt, Low Vt, High MV, Low MV, High Breath Rate, Low Breath Rate, Apnea, Disconnection, Rebreathing, High FiO2, Low FiO2, High SpO2, Low SpO2, High EtCO2, Low EtCO2, High InspCO2, High Pulse Rate, Low Pulse Rate, Low Last Power Source
Monitoring
- Displayed data
- Ppeak, PEEP, Pmean, Leakage, MV, Vt, FiO2, % in TgV, Total Rate, Spont Rate, % Spont, SpO2, Pulse Rate, I:E, Inspiratory Time, Rise Time, EtCO2, InspCO2
- Waveforms
- Pressure, Flow, Volume
- Trends over 1, 6, 24 and 48h
- Ppeak, PEEP, Total Rate, Spont Rate, Vt, Leakage, SpO2, EtCO2
Dimensions: W x H x D: 348 mm × 120 mm × 264 mm without click-on battery, (348 mm × 120 mm × 290 mm with click-on battery)
Weight: 5.2 kg without click-on battery (6.7 kg with click-on battery)
Noise level (at 1m, 10 cmH2O constant pressure): Less than 30 dB(A)